80 LPM Veterinary Oxygen System — For 24/7 Emergency Hospitals
42–80 LPM estimated demand · confirm operating reserve. Select the smallest oxygen plant that covers the estimated demand, then ask ON2 to confirm simultaneous use, operating reserve and site requirements. These are selection guides, not minimum operating loads.
Size from equipment demand
Count anesthesia machines, standard ventilators, dental units, oxygen doors and kennels at 3 LPM each; oxygen cages at 20 LPM and high-flow ventilators at 60 LPM. Confirm manufacturer-stated oxygen flow for specialized equipment.
Site preparation
Aura 10/20 use 120V plug-and-play connections. Aura 40 requires a dedicated 208V circuit. Aura 80 uses 208–240V, 32A single-phase with a 40A breaker and NEMA 14-50R. ON2 confirms model-specific room size and ventilation.
Service and timing
Annual preventive maintenance is included in the 5-year warranty. Typical lead time is 8–12 weeks after paperwork is complete.
Key benefits
- Highest capacity for 24/7 emergency and specialty hospitals
- Supports many concurrent stations plus sustained ICU oxygen
- Duplex redundancy available for uninterrupted supply
- Fixed monthly cost — never a surge oxygen invoice
Frequently asked questions
Is 80 LPM enough for a busy ER?
The 80 LPM system is the highest-capacity single unit and is designed for 24/7 emergency hospitals. For facilities that require guaranteed uninterrupted supply, ON2 pairs it with duplex redundancy.
How does on-site generation handle demand spikes?
The plant produces oxygen continuously and supply scales with draw, so sudden emergency demand does not leave you swapping cylinders mid-crisis.
Can we guarantee supply during maintenance?
Yes. A duplex configuration keeps oxygen flowing during service intervals, which is why most 24/7 hospitals choose redundancy at this capacity.
Build always-on oxygen for your ER
Use the calculator for a preliminary system and price estimate, or ask ON2 to review your clinic’s requirements.